IV set
|
PCV with automatic free flow protection; DEHP-free
|
|
Infusion Fluid
|
No restrictions, blood also acceptable
|
|
Flow rate
|
0.1 - 1200 ml/h,0.1ml/h increments up to 999.09 ml/h and above in Iml/h, priming the line: 1200m/h
|
|
Infusion Volume
|
0.1 - 20000 ml,0.1 ml increments
|
|
Bolus
|
Up t01200 ml/h,O.1 ml/h increments up to 999.9 ml/h and above in 1 ml/h;blous volume:1-500 ml Programmable volume (dose), time or flow rate. automatic and manual bolus;
|
|
KVO
|
O -20.0 ml/h
|
|
Flow rate accuracy
|
±5% over 96h or 10 liters, according to IEC 60601-2-24
|
|
Occlusion pressure
|
Downstream:11 levels :0.1-0.2 bar (75-900 mmHg)Pressure measurement accuracy +/-15% but not lower than 0.1 bar Upstream: automatic reduction of occlusion bolusGraphiv presentation of infusion pressure and set level (pictograms)
|
|
Drop Detector
|
Standard equipment, detachable
|
|
Air Detector
|
Ultrasound, programmable sensitivity: -single bubble: 0.05+0.3 ml -accumulated:O.I+I.5 ml/ 15 min
|
|
Operation Modes
|
ml/min, ml/h, ml/24h,l/h,l/24h, volume over time-automatic flow rate calculation Optional: ng,ml,ml+h, ug, mg,g, IAU, mU, U, kU, nmol, umol, mmol, mol/kg/min, h, 24h intermittent infusion profile-up to 24 cycles soft and hard limits for all infusion parameters
|
|
Additional options
|
Automatic free flow protectionDrug library up to 120 programmable drug dosing protocolsSoft and hard limits for all infusion parametersProgrammable infusion pause: 1 min-24hDelayed start: 1 min-24hProgrammed parameters protected by passwordDrug library and configuration pump to pump transferAlarm volume adjustment and choice of sound tvpeNiqht mode: tuninq the displav's briqhtness and alarm loudness, proqrammable time of start/ endSimple connection/ disconnection to docking station (also durinq infusion) Advanced pump testing systemPossibility to connect to the hospital information system directly or through docking station Open system: software update through RS232C
|
|
Induction Dose
|
Flow rate: up to 1200 ml/h, 0.1 ml/h increments; Programmable volume (dose), time or flow rate
|